Instruction Manual - TORBAL AGCN Series Torbal Analytical BalancesTable of ContentsChapter 1.Cautionary Notes and precaution...
Instruction Manual - TORBAL AGCN SeriesChapter 7.WEIGHING1. To begin weighing, press the power button ( ) to turn the scale ON. The scale will go th
Instruction Manual - TORBAL AGCN Series3. Once the stabilization indicator appears on the display, the container is ready to be tared. To tare the con
Instruction Manual - TORBAL AGCN SeriesChapter 8.CALIBRATIONWhen the scale is initially installed, it will automatically calibrate at start up to ensu
Instruction Manual - TORBAL AGCN Series4. The scale will display “WAIT”, indicating that automatic calibration has begun.5. Calibration can last from
Instruction Manual - TORBAL AGCN Series4. Calibration mass value settings which can be used for calibration adjustment will be displayed sequentially.
Instruction Manual - TORBAL AGCN Series9. Once the new calibration value is stored, the command “UNLOAD” will be displayed, indicating that this calib
Instruction Manual - TORBAL AGCN Series3. Press the T Key again when “CAL TM” is displayed. 4. Time interval settings will begin to display sequentia
Instruction Manual - TORBAL AGCN Series3. Press the T Key again when “CAL °C” is displayed.4. Temperature interval settings will begin to display sequ
Instruction Manual - TORBAL AGCN Series3. Press the T key again when command “CAL PRN” is displayed.4. Command “PRINT” will be displayed indicating th
Instruction Manual - TORBAL AGCN Series2. When command “AUTOTAR” is displayed, press the T key.3. Commands “AUT ON”,” AUT OFF”, and “AUT SET” will be
Instruction Manual - TORBAL AGCN Series9.1.2.ConfigurationThe AZSM range can be configured to work between 0.1d and 5d. To change the AZSM range, fol
Instruction Manual - TORBAL AGCN Series6. After the T key is pressed, in order to set a new AZSM range a dash will be displayed indicating a new AZSM
Instruction Manual - TORBAL AGCN SeriesChapter 10.UNITS OF MEASUREThe AGCN Series Scales can operate in three different units of measure: Grams (g), C
Instruction Manual - TORBAL AGCN SeriesChapter 11.PARTS COUNTING11.1.1.Using Default Sample Size Settings1. Place a container on the scale’s pan and p
Instruction Manual - TORBAL AGCN Series5. When “PCS” is displayed, press the T key. 6. Commands “PCS OFF” and “PCS ON” will be displayed sequentiall
Instruction Manual - TORBAL AGCN Series11.1.2.Using Custom Sample Size1. Place a container on the scale’s pan and press the T key to tare. After tar
Instruction Manual - TORBAL AGCN Series5. Commands “PCS OFF” and “PCS ON” will be displayed sequentially. To proceed with parts counting, press the
Instruction Manual - TORBAL AGCN Series8. Once your custom sample size value has been entered and the MENU key pressed, the scale will display the cou
Instruction Manual - TORBAL AGCN Series4. Commands “PCS OFF” and “PCS ON” will be displayed sequentially. To proceed with parts counting, press the
Instruction Manual - TORBAL AGCN Series7. Once the individual piece weight is entered and the MENU key pressed, the scale will display “0pcs”. At any
Instruction Manual - TORBAL AGCN SeriesChapter 1.CAUTIONARY NOTES AND PRECAUTIONThe TORBAL Precision scale is a sensitive and a delicate instrument. A
Instruction Manual - TORBAL AGCN Series2. Once the scale has been tared, place the sample reference weight on the pan. 3. The weight of the sample wi
Instruction Manual - TORBAL AGCN Series7. Remove the sample and place an object in the container. The scale will express the weight of that object a
Instruction Manual - TORBAL AGCN Series3. When “RECIPE” is displayed, press the T key. 4. Commands “REC OFF” and “REC ON” will be displayed sequenti
Instruction Manual - TORBAL AGCN Series9. Place a second ingredient in the container, wait until the weight stabilizes, and press the T key to tare.
Instruction Manual - TORBAL AGCN SeriesChapter 14.DENSITY CALCULATIONThe density calculation feature is used to assist in calculation of density in so
Instruction Manual - TORBAL AGCN Series14.1.1.SolidTo calculate the density of a solid, follow these steps:1. Press the M key to enter the main menu.2
Instruction Manual - TORBAL AGCN Series6. Commands H2O (distilled water), ETHANOL, and OTHER will scroll sequentially. Select the liquid that will b
Instruction Manual - TORBAL AGCN Series8. Once the temperature has been entered and accepted, command “PRESS TARE” will be displayed indicating to tar
Instruction Manual - TORBAL AGCN Series12. Once the weight of the object stabilizes, press the P key. Command “M2” followed by command “RESULT” will
Instruction Manual - TORBAL AGCN Series2. When command “DENSITY” is displayed, press the T key.3. Command “DEN ON” and “DEN OFF” will be displayed seq
Instruction Manual - TORBAL AGCN SeriesChapter 2.SPECIFICATIONS4ModelAGCN100 AGCN200Capacity100g 200gReadability (d)0.0001Repeatability (Standard Divi
Instruction Manual - TORBAL AGCN Series7. To key in the volume of the plunger, use the following keys: the C key to increment a digit, the T key to ac
Instruction Manual - TORBAL AGCN Series11. Submerge the plunger, by placing the beaker with the liquid on the base of the density apparatus.12. As the
Instruction Manual - TORBAL AGCN SeriesChapter 15.PRINTING AND RS232 COMMUNICATION PORT CONFIGURATIONAfter a weighing transaction is completed, a resu
Instruction Manual - TORBAL AGCN Series15.1.2.Baud1. Enter the main menu by pressing the MENU key.2. When “RS-232” is displayed, press the T key.3. Co
Instruction Manual - TORBAL AGCN Series5. The scale will begin to display available baud rates: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200. T
Instruction Manual - TORBAL AGCN Series4. Press the T key when “PARITY” is displayed.5. The scale will display available “PARITY” settings: “NONE”, “O
Instruction Manual - TORBAL AGCN Series3. Commands “BAUD”, “BITS”, and “PARITY” will begin to scroll sequentially.4. Press the T key when “BITS” is di
Instruction Manual - TORBAL AGCN SeriesChapter 16.DATE AND TIMETo set the current time and date, follow these steps.1. Enter the main menu by pressing
Instruction Manual - TORBAL AGCN Series2. The current time setting and the command “OUT” will scroll sequentially. To adjust the time, press T when
Instruction Manual - TORBAL AGCN Series16.1.2.Date1. To set the date, press T when “DATE” is displayed.2. The current date setting and the command “OU
Instruction Manual - TORBAL AGCN SeriesChapter 3.KEYS AND DISPLAY INDICATORSKey Primary Function Secondary FunctionØPower On and Power Off -T (YES)Tar
Instruction Manual - TORBAL AGCN Series6. Use the following keys to enter the current month setting: the C key to increment a digit, the T key to acce
Instruction Manual - TORBAL AGCN SeriesChapter 17.MENU CUSTOMIZATIONThe main menu can be customized to deactivate commands and functions that are not
Instruction Manual - TORBAL AGCN SeriesChapter 18.COMMON ERRORS AND TROUBLESHOOTINGError or Indicator Cause Explanation / Solution - - - - Below zero
Instruction Manual - TORBAL AGCN SeriesChapter 19.MAINTENANCECleaning and maintaining your AGCN Precision Scale:• Before cleaning the scale always unp
Instruction Manual - TORBAL AGCN SeriesChapter 22.LIMITED WARRANTYPURCHASER’S 24-MONTH WARRANTYWarranty is valid only if the attached warranty registr
Instruction Manual - TORBAL AGCN SeriesChapter 4.COMMANDS AND ABBREVIATIONSCommand or Abbreviation Description- - - - - -Taring in progressCALThe scal
Instruction Manual - TORBAL AGCN SeriesChapter 5.PARTS DESCRIPTION1 Pan 2 Pan Base3 Draft Ring4 Weighing Chamber Base5 Keypad6 Adjustable Legs7 Level
Instruction Manual - TORBAL AGCN SeriesChapter 6.UNPACKING THE BALANCE AND GETTING STARTED1. Carefully remove the scale, pan, and all of its component
Instruction Manual - TORBAL AGCN Series3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator shows